RHYTHMX ECG ANALYSIS PERFORMANCE
The AED RHYTHMx ECG Analysis system analyzes the patient’s ECG and advises you when the AED
detects a shockable or non-shockable rhythm.
This system makes it possible for a person, with no training in the interpretation of ECG rhythms, to offer
defibrillation therapy to victims of sudden cardiac arrest.
CARDIAC RHYTHMS USED TO TEST THE RHYTHM RECOGNITION DETECTION SYSTEM
FOR CARDIAC SCIENCE AED
Rhythm Class
Specifications
Shockable Rhythm – VF
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity
a
of >90%
Shockable Rhythm – VT
Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity
of >75%
Non-Shockable Rhythm –
NSR
Meets AAMI DF 39 requirement (>95%) and AHA recommendation
(>99%) of Specificity
Non-Shockable – Asystole
Meets AAMI DF 39 requirement and AHA recommendation of Specificity
of >95%
Non-Shockable– all other
rhythms
Meets AAMI DF 39 requirement and AHA recommendation of Specificity
of >95%
a
Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting
Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms and Enhancing Safety, AHA AED Task Force
and approved by the AHA Science Advisory and Coordinating Committee. Circulation, 1997(95), pp 1677-1682