SECTION 3 – MAINTENANCE AND TROUBLESHOOTING
Page 14 of 35
© 2010 General Electric Company
2022100-201 Rev C
RHYTHMX® AED ECG ANALYSIS ALGORHITHM
The RhythmX AED ECG analysis algorithm provides superior ECG detection capabilities, allowing it to be
placed on patients at risk for sudden cardiac arrest. The features available with the AED include the
following:
• Detection
Rate
• Asystole
Threshold
• Noise
Detection
• Non-Committed
Shock
• Synchronized
Shock
•
Pacemaker Pulse Rejection
• SVT
Discriminators
•
Supraventricular Tachycardia (SVT) Rate
• Continuous
Monitoring
DETECTION RATE
All ventricular fibrillation (VF) and ventricular tachycardia (VT) rhythms at or above this rate will be classified
as shockable. All rhythms below this rate will be classified as non-shockable. This rate is programmable
between 120 bpm (beats per minute) and 240 bpm via ServiceLink Software by the Medical Director. The
default Detection Rate is 160 bpm. The Responder AED detection rate is 160 bpm.
ASYSTOLE THRESHOLD
The asystole baseline-to-peak threshold is set at 0.08 mV. ECG rhythms at or below 0.08 mV will be
classified as Asystole and will not be shockable.
NOISE DETECTION
The AED will detect noise artifact in the ECG. Noise could be introduced by excessive moving of the patient
or electronic noise from external sources like cellular and radiotelephones. When noise is detected, the AED
will issue the prompt “
ANALYSIS INTERRUPTED. STOP PATIENT MOTION”
to warn the operator. The
AED will then proceed to reanalyze the rhythm and continue with the rescue.
NON-COMMITTED SHOCK
After the AED advises a shock, it continues to monitor the patient ECG rhythm. If the patient’s rhythm
changes to a non-shockable rhythm before the actual shock is delivered, the AED will advise that the rhythm
has changed and issue the prompt “
RHYTHM CHANGED. SHOCK CANCELLED
.” The AED will override
the charge and continue ECG analysis.
SYNCHRONIZED SHOCK
The AED is designed to synchronize shock delivery on the R-wave. The AED will automatically attempt to
synchronize to the R-wave. If delivery cannot be synchronized within one second, a non-synchronized shock
will be delivered.