SECTION 4 – TECHNICAL DATA
Page 32 of 35
© 2010 General Electric Company
2022100-201 Rev C
Figure A1. STAR BIPHASIC WAVEFORM
Table A1 - Ultra-Low Variable Energy (150 VE) Responder AED Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(ms)
Voltage*
(Volts)
Duration*
(ms)
Energy**
(Joules)
25
1393
3.3
743
3.2
145-196
50
1420
4.5
909
3.2
128-173
75
1430
5.8
973
3.2
116-156
100
1434
7.0
1007
3.2
108-146
125
1437
8.3
1027
3.2
102-138
150 1439 9.5 1040 3.2
98-132
175 1441 10.8 1049 3.2
95-128
Table A2 – Low Variable Energy (200 VE) Waveform Responder AED Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(ms)
Voltage*
(Volts)
Duration*
(ms)
Energy**
(Joules)
25
1609
3.3
858
3.2
193-260
50
1640
4.5
1050
3.2
170-230
75
1651
5.8
1124
3.2
155-209
100
1656
7.0
1163
3.2
144-194
125
1660
8.3
1186
3.2
136-184
150 1662 9.5 1201 3.2 131-176
175 1663 10.8 1212 3.2 126-170