SECTION 4 – TECHNICAL DATA
Page 34 of 35
© 2010 General Electric Company
2022100-201 Rev C
These protocols are selected by using our ServiceLink software program. The five biphasic energy protocols
available are as follows:
Energy Protocols
Shock Sequence
1
Energy Level (VE)
Energy Range
2
(J)
Factory Default
1.
200
126-260
2.
300
170-351
3.
300
170-351
Protocol #2
1.
200
126-260
2.
200
126-260
3.
300
170-351
Protocol #3
1.
150
95-196
2.
200
126-260
3.
200
126-260
Protocol #4
1.
150
95-196
2.
150
95-196
3.
200
126-260
Protocol #5
1.
200
126-260
2.
200
126-260
3.
200
126-260
1
The Ultra-Low Energy (150 VE), Low Energy (200 VE) and High Energy (300 VE) shocks are variable energy. The
actual energy is determined by the patient’s impedance.
2
Allowable energy range