SECTION 3 – MAINTENANCE AND TROUBLESHOOTING
2022100-201 Rev C
© 2010 General Electric Company
Page 19 of 35
PADS
The defibrillation pads come in a ready-to-use, sealed package containing one pair of
self-adhesive pads with an attached cable and connector. The pads are disposable
and should be discarded after each rescue. The pads have a limited shelf life and
shall not be used beyond the expiration date. Keep a fresh, unopened pair of pads
plugged into the AED at all times. Refer to the electrode package label for operation
temperatures.
On the Responder AED, an audible and visual alert will indicate after the self-test if the pads are missing, unplugged or
damaged.
CAUTION:
Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance. Reuse of pads may result in improper
AED performance
.
CAUTION:
Do not reuse pads – Used pads may not adhere properly to patient.
Improper pad adhesion may result in skin burns
.
Improper pad adhesion may result in improper AED performance.
Used pads may cause patient to patient contamination
.
ELECTRODE INSTALLATION
1. Open the lid of the AED.
2. Place the package into the lid so that the expiration label is visible
through the clear window on the lid. The expiration date of the pads will
then be readable without opening the lid of the AED.
3. Match the color of the connectors (red to red), then plug the electrode
connector into the AED case as shown in the drawing.
4. Tuck the excess cable length in the bottom holder as shown in the
drawing. With the package completely secured to the AED lid, close the
lid.
5. Make sure the expiration date is visible through the clear window of the
lid. Make sure that the
STATUS INDICATOR
is
GREEN
.
CAUTION:
Use only Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by
General Electric may cause the AED to function improperly during a rescue.
CAUTION:
Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance.
DIRECTIONS FOR USE:
1.
Do NOT open until ready to use, short term use only.
2.
Ensure the skin site is clean and dry.
3.
Separate one pad from liner.
4.
Place one pad on skin in either location.
5.
Peel and place remaining pad.