38
Service Instruction
───────────────────────────────────────────────────────────────────
227 487 20 Rev G
type:
TRIO MATE 4pol
signal
name level
active
I/O
meaning
1 SHOTAFOL_
+5V Low Input key
“shock”
2 CHATAFOL_
+5V Low Input key
“charge/shock
3 SYNTAFOL_
+5V Low Input key
“sync”
4 GND
0
-
Output
Ground
connector :
PaddConn1
function:
connector to “Sternum” Paddle
type:
MODU II 4pol
signal
name level
active
I/O
meaning
1
SHOTAPAD
+5V
High
Input
key “shock” (paddle)
2 VCC
+5V
-
Output
VCC
3 PADTYP0
+5V High Input paddle
detection
4 PADTYPHE2
analog
Input paddle
detection (level depends on paddle type)
connector :
PaddConn2
function:
connector to “Apex” paddle
type:
MODU II 4pol
signal
name level
active
I/O
meaning
1 NC
2 NC
3
CHATAPAD
+5V
High
Input
key charge/shock (paddle)
4 VCC
+5V
-
Output
VCC
connector :
PaddleHVConn
function:
connector to paddles (high voltage)
type:
4,8 x 0,8
signal
name level
active
I/O
meaning
7
HV_POS
5kV
-
Output
connection to “Sternum” paddle
8
HV_NEG
0V
-
Output
connection to “Apex” paddle
Содержание Responder 1000
Страница 1: ...Responder 1000 1100 Version V 1 0 Servicing Instructions 227 487 20 ENG Revision G ...
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 53: ......
Страница 54: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......