GE Medical Systems
Information Technologies
Responder 1000/1100
Page 9
/
38
Service Instruction
───────────────────────────────────────────────────────────────────
2.2 Responder 1100
1
ECG trigger signal input (synchronized defibrillation)
2
Button to deliver the shock - together with button
12
3
AC line power indicator
4
Button to enable and disable the synchronized defibrillation mode, with indicator (indicator
is illuminated during sync. defibrillation and flashes to the rhythm of the QRS complex)
5
Indicator light is green while the battery is charging (line power operation), it is red when
the battery needs charging (battery power operation)
6
Energy selector, On/Off switch
7
Indicators
2
to
360
illuminate when the corresponding energy level has been reached
8
"On" indicator illuminates when the unit is On
9
"Service" indicator illuminates when a problem was detected during the automatic self-test
10
Button to initiate defibrillator charging and to deliver the shock - together with button
11
when adhesive pads or internal electrodes are used
11
Button to deliver the shock when adhesive pads or internal electrodes are used - together
with button
10
12
Button to initiate defibrillator charging and to deliver the shock - together with button
2
13
Power
cord
14
Tube holder for skin prep cream
15
Connector for defibrillation electrodes (turn off the unit before exchanging the electrodes)
227 487 20 Rev G
Содержание Responder 1000
Страница 1: ...Responder 1000 1100 Version V 1 0 Servicing Instructions 227 487 20 ENG Revision G ...
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 53: ......
Страница 54: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......