Advanced Programming
3-43
March 2008
Entering Oxygen Curve
Data
Press the
[NO]
key to return to the
Select O2 Curve Points
menu. Use
the arrow keys to select SPAN and enter the span microamp (IO1),
and ppmv or % (IO1) values.
You may now do one of the following:
•
Enter data for a different channel or type of sensor by pressing the
[NO]
key until you reach the Programming Menu, and then refer to
Entering Calibration Data for New Probes/Sensors
on page 3-39.
•
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
•
Exit by pressing the
[RUN]
key.
IMPORTANT:
You must configure a channel AND enter the
necessary calibration data in order for a sensor to
work properly.
O2 Curve Menu 1
Use the arrow keys to move the
brackets to
S/N
and press
[YES]
.
[S/N] CURVE BkGd
Probe Serial # 1
Enter the serial number and press
[YES]
. Refer to the Calibration
Data sheet for the serial number.
Press
[NO]
to return to the Curve
Menu.
O2: 123456
O2 Curve Menu 1
Use the arrow keys to move the
brackets to
CURVE
and press
[YES]
.
S/N [CURVE] BkGd
Sel. O2 Curve Pts#
Use the arrow keys to move the
brackets to
ZERO
and press
[ZERO] SPAN
[YES].
#1 O(uA) O(ppm)
Enter the zero microamp value
(IO0) that is shown on the
Calibration Data Sheet. Press
[YES]
and press the right arrow
key.
0.721 0.0000
#1 O(uA) O(ppm)
Enter the zero ppmv (or %) value
(OX0) and press
[YES]
.
0.721 0.5000