Advanced Programming
3-35
March 2008
Reconfiguring a Channel
for a New Sensor
If you decide to connect a different type of sensor to one of the
channels of your unit, you must reconfigure that channel. For
example, if you are using a moisture probe on channel 1 and later
decide to connect an oxygen cell, you must “tell” (reconfigure) the
MMS-3
that you have changed the sensor.
Note:
If you are using a probe capable of multiple inputs (e.g., TF
probe) or using more than one input (including a constant),
you must configure the channel to measure all the possible
inputs. Refer to the
Multiple Input Instructions
highlighted in
this procedure.
Note:
If you are already in the System Menu, proceed to the
Measurement Mode
step.
Press the
[PROG]
key to enter the user program.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
[CHAN]
key to select the desired channel.
Enter Passcode: XXXX
Enter the passcode
.
Programming Menu 1
Use the arrow keys to move the
brackets to
SYSTEM
and press
[YES].
[SYSTEM] AUTOCAL
`
Measurement Mode 1
Select the desired measurement
mode and press
[YES].
See
Table 3-9 below for a list of
available measurement modes.
O [H] T P Aux1
`
Table 3-9: Measurement Modes
Display Abbreviation
Measurement Mode
O
Oxygen
H
Hygrometry
T
Temperature
P
Pressure
AUX1
Auxiliary 1
AUX2
Auxiliary 2
CNST-PPMV
PPMv Multiplication Factor