Advanced Programming
3-3
March 2008
Setting Up the Alarms
(cont.)
The deadband allows users to set a programmable range of (normally
small) values beyond the setpoint, so that the
MMS-3
will not trigger
an alarm unless the measured value goes outside the deadband range.
On a High (Above) alarm, the alarm will trip if the measurement is
greater than or equal to the setpoint. It will not
RESET
until the
measurement is less than the setpoint
minus
the deadband. However,
on a Low (Below) alarm, the alarm will trip if the measurement is less
than or equal to the setpoint. It will not
RESET
until the measurement
is greater than the setpoint
plus
the deadband.
You may now do one of the following:
•
Set up the alarms for the other channel (page 3-1). Begin by
checking the channel number.
•
Refer to another section and perform a different procedure. Refer
to the menu maps at the end of this chapter to navigate through the
user program.
•
Exit by pressing the
[RUN]
key.