Advanced Programming
3-31
March 2008
Using Computer
Enhanced Response
(cont.)
Press the
[RUN]
key and wait for the
MMS-3
to begin displaying
measurements. Abruptly expose the sensor to dry gas.
IMPORTANT:
You must enable Computer Enhanced Response
before you expose the sensor to dry gas. The
MMS-3
must see all the dry down in order to make an
accurate calculation.
When you activate Computer Enhanced Response for a channel, an
EH replaces the H as the measurement mode. The
MMS-3
will
display a reverse video E next to the channel number during dry
down. The reverse video E will change to a regular E when the
MMS-3
determines the final value.
You may now do one of the following:
•
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
•
Exit by pressing the
[R
UN]
key.
Enhanced Response 1
Use the arrow keys to move the
brackets to the desired selection
and press
[YES].
NO [YES]
Содержание Moisture monitor series 3