Chapter 10 - Care & Maintenance
10-33
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Type BF Patient Leakage Current Test - Mains to Probe (This test should be done for all probes)
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
Table 10-23
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.1mA
Closed
Normal
Open
0.5mA
Closed
Reversed
Closed
0.1mA
Closed
Reversed
Open
0.5mA
Open
Normal
Closed
0.5mA
Open
Reversed
Closed
0.5mA
Table 10-24
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
5mA
Closed
Reversed
Closed
5mA