Chapter 10 - Care & Maintenance
10-7
10-5-2
Functional Checks
The functional checks take about 60 minutes to perform.
Refer to the LOGIQ™ S7 Expert/Pro Basic User Manual whenever necessary.
10-5-2-1
System Checks
NOTE:
* Some software may be considered standard depending upon system configuration.
Table 10-5 System Functional Checks
Step
Item
Description
1
B Mode
Verify basic B Mode (2D) operation.
Check the basic system controls that affect this mode of operation.
2
M Mode
Verify basic M Mode operation.
Check the basic system controls that affect this mode of operation.
3
CFM Mode
Verify basic CFM Mode (Color Flow Mode) operation.
Check the basic system controls that affect this mode of operation.
4
PD Mode
Verify basic PD Mode (Power Doppler Mode) operation.
Check the basic system controls that affect this mode of operation.
5
Doppler Modes
Verify basic Doppler Mode operation (PW ).
Check the basic system controls that affect this mode of operation.
6
*Applicable Software
Options
Verify the basic operation of all optional modes.
Check the basic system controls that affect each options operation.
7
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
8
LCD Monitor
Verify basic LCD Monitor display functions.
9
Measurements
Scan a gray scale phantom and use the measurement controls to verify distance and area
calculation accuracy.
Refer to the Basic User Manual, for measurement accuracy specifications.