Chapter 4 - Functional Checks
4-23
4-3-6-6
Gel Warmer Test
1.) Insert Gel Bottle into Gel Warmer.
2.) Power ON the warmer.
3.) Verify Gel gets warm.
NOTE:
Depending on environment, Gel requres 30 minute or more to warm up.
4-3-6-7
Single CWD
1.) Connect P2D or P6D probe. Select cardiac application like “Adult”. And start scan.
2.) Verify that CW timeline image appears.
4-3-7
Mode Transition Checks
4-3-7-1
General Information
Figure 4-21
4-3-7-2
B-Mode
1.) Press B Mode button.
2.) Verify that system scan with the activated probe.
3.) Verify probe name is displayed on upper right hand corner of main monitor (scan screen).
B-Flow
B
CF
PW
M