10-8
Section 10-5 - System Maintenance
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
10-5-3
Input Power
10-5-3-1
Mains Cable Inspection
10-5-4
Cleaning
10-5-4-1
General Cleaning
Frequent and diligent cleaning of the LOGIQ™ S7 ultrasound system reduces the risk of spreading
infection from person to person, and also helps to maintain a clean work environment.
Table 10-6 Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
2
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
3
Color Deskjet
(Bluetooth) Printer
Verify hardcopy output of the Deskjet (Bluetooth) printer. Clean heads and covers if necessary.
4
DVR(R1.x.x)
Verify record/playback capabilities of the DVR.
5
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
6
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
7
DVD-Drive
Verify that the DVD-drive reads/writes properly
(export/recall image in Image Management System = Archive)
Table 10-7 Mains Cable Inspection
Step
Item
Description
1
Unplug Cord
Disconnect the mains cable from the wall and system.
2
Inspect
Inspect it and its connectors for damage of any kind.
3
Terminals
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4
Inlet Connector Inlet connector retainer is functional.
Table 10-8 General Cleaning
Step
Item
Description
1
Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
Caution:
DO NOT allow any liquid to drip or seep into the system.
2
LCD Monitor
Clean LCD Monitor surface with a fluid detergent in warm water on a soft, damp cloth.
Caution:
DO NOT spray any liquid directly onto the LOGIQ™ S7 Expert/Pro covers, LCD Monitor,
keyboard, etc.
Содержание LOGIQ S7 Expert