GE H
EALTHCARE
DRAFT
LOGIQ™ S7 E
XPERT
/P
RO
D
IRECTION
5460683
, R
EVISION
3
DRAFT (J
UNE
25, 2014)
S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-9
7-4-3
System Warning/Errors and Logs
7-4-3-1
Temperature warning
7-4-3-1-1
Cause
CPU Temperature exceeded threshold (95°C).
7-4-3-1-2
What to do
1.Shut down the system and Clean the filter.
7-4-3-2
Temperature exceeded threshold
7-4-3-2-1
Cause
CPU Temperature exceeded threshold (100°C) and therefore, system must be shut down immediately.
7-4-3-2-2
What to do
1.Shut down the system and Clean the filter.
7-4-3-2-3
Log message of the high temperature
Example
Error ; acqHouseKeeping(****); CPU Temperature Sensor BEP_CPU_TEMPERATURE = 97.00
degrees C is outside tolerance limits. UpperLimit = 100.00 LowerLimit = -10.00
Содержание LOGIQ S7 Expert
Страница 2: ......
Страница 11: ...GE HEALTHCARE LOGIQ S7 EXPERT PRO DIRECTION 5460683 REVISION 3 SERVICE MANUAL ix ZH CN KO ...
Страница 38: ...GE HEALTHCARE LOGIQ S7 EXPERT PRO DIRECTION 5460683 REVISION 3 SERVICE MANUAL xxxvi Table of Contents ...
Страница 434: ......