Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 10 Care & Maintenance
10-15
10-6-4-3
Data Sheet for enclosure Source Leakage Current
The test passes when all readings measure less than the value shown in
. Record all data
on the PM Inspection Certificate.
Table 10-14 Typical Data Sheet for enclosure Source Leakage Current
Unit Power
Tester Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1
Speaker
Cover
Test 2
Real Panel
Metal
Parts
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
ON
NORM
OPEN
ON
NORM
CLOSED
ON
REV
OPEN
ON
REV
CLOSED
OFF
NORM
OPEN
OFF
NORM
CLOSED
OFF
REV
OPEN
OFF
REV
CLOSED