GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
8-2
Section 8-2 - Disassembly/Re-assembly of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
8-2-2
Handle Assy (FRU No. 313)
Purpose: This is a description on how to remove and replace the Handle Assy (FRU No.313).
8-2-2-1
Tools
•
Plier
8-2-2-2
Needed Manpower
•
1 person, 2 m travel
8-2-2-3
Preparations
•
Shut down the system, then disconnect the AC/DC power cord and remove the battery.
8-2-2-4
Removal Procedure
1.) Unscrew the two handle caps on both sides of the system, the rotation direction is
counterclockwise.
2.) Pull out the Handle.
8-2-2-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.