GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 3 System Setup
3-27
L.) Connect the external monitor to the system. The monitor can be properly connected using the VGA
port.
Press Ctrl+Alt+V on the keyboard, a dialog box appears.
Select Intel(R) Dual Display Clone; check the Same display configuration driver on both display box
and select OK.
NOTE:
Please refer to the operation manual of each peripheral for information needed by the user to
operate the system safely.
NOTE:
For LOGIQ e R5.2.x, R6.x.x, R7.x.x, LOGIQ e Vet R6.x.x, R7.x.x, the default set is dual, it will
not pop up Figure 3-28
For detailed installation information, please refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Peripheral Installation manual.
Figure 3-27 VGA Connection
Figure 3-28 Graphics Controller Properties
For LOGIQ e R5.x.x, R6.x.x, R7.x.x, LOGIQ i R5.x.x,
For LOGIQ e R4.x.x, LOGIQ i R4.x.x, Vivid e R4.x.x
Vivid e R5.x.x/R6.x.x, LOGIQ e Vet R6.x.x, R7.x.x