LOGIQ C5/C5 PRO Quick Guide
Direction 5272220-100 Rev. 2
47
Safety
Best practices while scanning
NOTE: Refer to the Optimization sections of the Modes chapter for a complete discussion of each control.
Acoustic Output Default Levels
In order to ensure that an exam does not start at a high output level, the LOGIQ C5/C5 PRO initiates scanning at a reduced default output level. This reduced level
is a programmable preset and depends upon the exam category and probe selected. It takes effect when the system is powered on or
New Patient
is selected.
HINTS
Increase the Acoustic Output only after attempting image optimization with controls that have no effect on Acoustic Output, such as Gain
and TGC.
WARNING
Be sure to have read and understood control explanations for each Mode used before attempting to adjust the Acoustic Output control or
any control that can effect Acoustic Output.
Acoustic
Output
Hazard
Use the minimum necessary acoustic output to get the best diagnostic image or measurement during an examination. Begin the exam with
the probe that provides an optimum focal depth and penetration.