LOGIQ C5/C5 PRO Quick Guide
Direction 5272220-100 Rev. 2
38
EMC Performance
All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, either transmitted through air or connecting
cables. The term EMC (Electromagnetic Compatibility) indicates the capability of equipment to curb electromagnetic influence from other equipment and at the same
time not affect other equipment with similar electromagnetic radiation from itself.
Proper installation following the service manual is required in order to achieve the full EMC performance of the product.
The product must be installed as stipulated
on page 39 ”Notice upon Installation of Product”
.
In case of issues related to EMC, please call your service personnel.
The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifications
to this equipment. Unauthorized changes or modifications could void the users’ authority to operate the equipment.
CAUTION
Do not use devices which intentionally transmit RF signals (cellular phones, transceivers, or radio controlled products) in the vicinity of this
equipment as it may cause performance outside the published specifications. Keep the power to these type devices turned off when near
this equipment.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be around this equipment
to fully comply with the above regulation.