LOGIQ C5/C5 PRO Quick Guide
Direction 5272220-100 Rev. 2
35
Classifications
Type of protection against electric shock
LOGIQ C5/C5 PRO console is Class I Equipment (*1)
Degree of protection against electric shock
Type BF Applied part (*3)
(for Probes marked with BF symbol)
Type CF Applied part (*4)
(for ECG marked with CF symbol)
Continuous Operation
System is Ordinary Equipment (IPX0)
Footswitch is IPX1 (for IPX1 is FSU-2001).
*1. Class I EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but includes an protective earth ground. This additional safety
precaution prevents exposed metal parts from becoming LIVE in the event of an insulation failure.
*2. Class Il EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but in which additional safety precautions such as DOUBLE
INSULATION or REINFORCED INSULATION are provided.
*3. Type BF APPLIED PART
TYPE BF APPLIED PART providing a specified degree of protection against electric shock, with particular regard to allowable LEAKAGE CURRENT.
Table 7: Type BF Equipment
Normal Mode
Single fault condition
Patient leakage current
Less than 100 microA
Less than 500 microA