LOGIQ 7/LOGIQ 7 Pro Quick Guide
Direction 5307393-100 Rev. 1
13
Volume
1.
To make a volume calculation, do one of the
following:
•
Make one distance measurement.
•
Make two distance measurements.
•
Make three distance measurements.
NOTE: Three distances should be done in the dual
format mode (side by side images). One
measurement is usually made in the sagittal plane
and two measurements in the axial plane.
•
Make one distance and one ellipse measurement.
•
Make one ellipse measurement.
2.
Select
Volume
.
Time Interval Measurement
1.
Press
Measure
twice; and active caliper with a
vertical dotted line displays.
2.
To position the active caliper at the start point, move
the
Trackball
.
3.
To fix the start point, press
Set
. The system fixes the
first caliper and displays a second active caliper.
4.
To position the second caliper at the end point, move
the
Trackball
.
5.
To complete the measurement, press
Set
. The
system displays the time interval between the two
calipers in the measurement results window.
Velocity Measurement
1.
Press
Measure
; an active caliper with a vertical
dotted line displays.
2.
To position the caliper at the desired measurement
point, move the
Trackball
.
3.
To complete the measurement, press
Set
. The
system displays the velocity measurement in the
measurement results window.
PI, RI, PS/ED Ratio, ED/PS Ratio or A/B Ratio
Select
PI
,
RI
,
PS/ED Ratio
,
A/B Ratio
or
ED/PS Ratio
from the Doppler Touch Panel. Perform velocity
measurements.
1.
The first caliper is the start point on the Doppler
waveform. This would be V
MAX
for PI, peak velocity
for RI, systole for PS/ED ratio, “A” velocity for A/B
ratio or diastole for ED/PS ratio.
2.
The second caliper is the end-point caliper to the end
point of the Doppler waveform. This would be V
d
for
PI, minimum velocity for RI, diastole for PS/ED ratio,
“B” velocity for A/B ratio or systole for ED/PS ratio.
NOTE:
For the PI calculation, if Trace Auto is not
selected, manually trace the waveform between V
MAX
and V
d
.
NOTE:
For the PI calculation, if Trace Auto is on, the
system automatically traces the waveform when Set is
pressed to fix V
d
.
Worksheets
Measurement/Calculation worksheets are available to
display and edit measurements. There are generic
worksheets as well as Application specific worksheets.
The worksheets are selected from the Measurement
Touch Panel.
Reports
The system enables the generation of patient reports,
based on the examination performed and the analyses
that were made during examinations. The reports are
generated using the data stored in the system with pre-
selected templates.
For more information, see the LOGIQ 7/LOGIQ 7 Pro
Basic User Manual Chapter 14 “ReportWriter”.
Activating the report
1.
Select
Report
on the Touch Panel. Report page is
displayed.
2.
Select the desired application and template by
TEMPLATE
.
3.
Edit the report as necessary.
•
Change the Patient Data
•
Insert the images or anatomical graphics.
•
Type the examiner’s comment, etc.
4.
Print the report as necessary.
5.
Select Store to save the report. Select another key to
close the report.