LOGIQ 7/LOGIQ 7 Pro Quick Guide
Direction 5307393-100 Rev. 1
11
Basic Measurements
NOTE:
The following instructions assume that you first
scan the patient and then press
Freeze
.
Distance and Tissue Depth Measurements
1.
Press
Measure
once; an active caliper displays.
2.
To position the active caliper at the start point
(distance) or the most anterior point (tissue depth),
move the
Trackball
.
3.
To fix the start point, press
Set
. The system fixes the
first caliper and displays a second active caliper.
4.
To position the second active caliper at the end point
(distance) or the most posterior point (tissue depth),
move the
Trackball
.
5.
To complete the measurement, press
Set
. The
system displays the distance or tissue depth value in
the measurement results window.
NOTE:
Before you complete a measurement:
To toggle between active calipers, press
Measure
.
To erase the second caliper and the current data
measured and start the measurement again,
press
Clear
once.
NOTE:
To rotate through and activate previously fixed
calipers, turn
Cursor Select
.
NOTE:
After you complete the measurement, to erase
all data that has been measured to this point, but not data
entered onto worksheets, press
Clear
.
Circumference/Area (Ellipse) Measurement
1.
Press
Measure
once; an active caliper displays.
2.
To position the active caliper, move the
Trackball
.
3.
To fix the start point, press
Set
. The system fixes the
first caliper and displays a second active caliper.
4.
To position the second caliper, move the
Trackball
.
5.
Turn the
Ellipse
control; an ellipse with an initial
circle shape appears.
NOTE:
Be careful not to press the Ellipse control as this
activates the Body Pattern.
6.
To position the ellipse and to size the measured axes
(move the calipers), move the
Trackball
.
7.
To increase the size, turn the
Ellipse
control in a
clockwise direction. To decrease the size, turn the
Ellipse
control in a counterclockwise direction.
8.
To toggle between active calipers, press
Measure
.
9.
To complete the measurement, press
Set
. The
system displays the circumference and area in the
measurement results window.
NOTE:
Before you complete a measurement:
To erase the ellipse and the current data
measured, press
Clear
once. The original
caliper is displayed to restart the measurement.
To exit the measurement function without
completing the measurement, press
Clear
a
second time.
Circumference/Area (Trace) Measurement
1.
Press
Measure
twice; a trace caliper displays.
2.
To position the trace caliper at the start point, move
the
Trackball
.
3.
To fix the trace start point, press
Set
. The trace
caliper changes to an active caliper.
4.
To trace the measurement area, move the
Trackball
around the anatomy. A dotted line shows the traced
area.
NOTE:
To erase the dotted line but not the trace caliper,
press
Clear
once. To clear the trace caliper and the
current data measured, press
Clear
twice.
NOTE:
To erase the line (bit by bit) back from its current
point, move the
Trackball
or turn the
Ellipse
control
counterclockwise.
5.
To complete the measurement, press
Set
. The
system displays the circumference and the area in
the measurement results window.
NOTE:
Before you complete a measurement:
To erase the line (bit by bit) back from its current
point, move the Trackball or turn the Ellipse
control counterclockwise.
To erase the dotted line but not the trace caliper,
press Clear once.
To clear the trace caliper and the current data
measured, press Clear twice.