GE M
EDICAL
S
YSTEMS
D
IRECTION
2317229, R
EVISION
3
LOGIQ™ 180 B
ASIC
S
ERVICE
M
ANUAL
Chapter 3 Installation
3 - 21
3-6-2-3
Storage & Operation Requirements
The LOGIQ™ 180 is shipped in a single container excluding probes. Shipping weight is approximately
12 kgs. Table 3-5 provides a summary of temperature, atmospheric pressure, and humidity tolerances
for shipping, installation and operation:
Table 3-5
Storage & Operation Requirements
3-6-3
Optional Peripherals
LOGIQ™ 180 peripherals and accessories can be properly connected using the connectors on the rear
panel of the LOGIQ™ 180 system.
External Optional Peripherals
•
VCR (Sony SVO-9500MD)
•
VGP (Sony UP-895MDW/MD/SYN)
Parameter
Storage
Operation
Temperature (Deg.C)
(Deg. F)
-10 to 60
14 to 140
10-40
50 to 104
Atmospheric Pressure (hPa)
700 to 1060
700 to 1060
Humidity (%)
Non- Condensing
30 to 80
35 to 75
WARNING
WARNING
Parallel Printing should not be done while live scanning patients. Also, the Printer
cable should not be attached to the LOGIQ™ 180 during the scan.