Chapter 4 Functional Checks
4 - 11
Section 4-7
Patient contact tools
4-7-1
Probe/Connectors Check
Take the probes and check them as described below:.
Table 4-12 Probe and connectors check
Step
Check
Expected Result(s)
If Not Remedy
1.
To test each delivered Probe, Connect it to
the System
It will display the probe name on the
screen
2.
Hold the probe connector horizontally with
the cable pointing Sideways.
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port and
carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
Position the probe cable so that it is not
resting on the floor
CAUTION:
Do not allow the probe head to hang freely.
Impact to the probe head may result in
irreparable damage.
To connect a probe:
3.
Rotate the lock handle counter-clockwise
to the Vertical position to unlock the
connector.
Remove the connector from the port
Ensure that the probe head is clean before
placing the probe in its storage case.
To disconnect probes:
The probes that are not connected to
the unit should be stored in their
storage case.
4.
Two Probes can be Switched by pressing
the Key on the Two Probe Port
The LED toggles to indicate the probe
change & a click is also heard to indicate
change over
The system initalizes the new probe & the
image from the newly selected probe is
displayed in the B-Mode
5.
Presetting Parameters to a Probe, Select
the desired probe & adjust the desired
parameters
Press CTRL + W + Enter
The Parameters selected will be set as
Default Values