Chapter 5 Theory
5 - 5
5-2-1
PCB Nomenclature
The following table lists the Circuit boards on LOGIQ™ 180 system.
5-2-2
Dip Switch Setting
5-2-2-1
FEB Board
Table 5-14 Circuit board on LOGIQ™ 180
Board Abbrv
Description
Remarks
CPU
Central Processing Unit
For LOGIQ™
180 (V7.0HC)
DSC
Digital Scan Converter
For LOGIQ™
180 (V7.0HC)
KBD
Keyboard
For LOGIQ™
180 (V7.0HC)
FEB
Front End Board
For LOGIQ™
180 (V7.0HC)
PDB
Power Distributor PCB
For LOGIQ™
180 (V7.0HC)
RLB
Relay Board
For LOGIQ™
180 (V7.0HC)
CAUTION
The Settings are valid only when jumper JP1 and JP20 are in Test mode and is used for PCB
testing in the production line. The equipment at the customer site should have JP1 and JP20 in
normal mode which makes DIP switch S1 ineffective.
Table 5-15
S1
1
Probe Type 0
2
Probe Type 1
3
Probe Type 2
4
Probe Type 3
5
Probe Type 4
6
SDO (Shutdown 0)
0: Enables probe type indicator (Indicates probe is connected)
1: Disables probe type indication (Indicates probe is disconnected)
7
SD1 (Shutdown 1)
0: HV low (Not Used)
1: HV high (Not Used)