Krautkramer USM 35X
Issue 01, 04/2005
5-29
Operation
D = A – B – C
with:
• A = Indication (in dB)
Absolute instrument gain with which the maximum
flaw echo is at 50 % (±5 %) echo height
• B = Reference (in dB)
Absolute instrument gain with which the maximum
reference echo (1.5 mm side-drilled hole from the
reference block 1) is at 50 % (±5 %) echo height
•
C = Attenuation (in dB)
This value is calculated according to the formula
C = 0.079 dB/mm (s – 25.4 mm). With s = sound
path of the flaw echo.
The sound attenuation correction is automatically
calculated and displayed by the instrument. For
sound paths smaller than or equal to 25.4 mm
(1 inch), the value is set to zero.
• D = D 1.1 Rating (in dB)
This is the result of the evaluation according to AWS.
The evaluation is carried out in the USM 35X accord-
ing to the formula indicated above.
Classification of welds (function group AWS)
H
Notes:
Make sure that all instrument options for the special
test are calibrated before starting with the rating ac-
cording to AWS.
Pay attention to peaking an echo with an amplitude
between 45 % and 55 % screen height. A rating is not
possible with other amplitudes.
– Apply couplant, and couple the probe to the refer-
ence block 1. Peak the echo from the 1.5 mm side-
drilled hole.
– Select the function
aSTART
, and set up the A gate
on the reference echo.
– Vary the gain so that the reference echo is displayed
at 50 % screen height.
– Choose the function
REFRNCE
, and confirm the
choice in order to save the reference gain.
Krautkramer USM 35X
Issue 01, 04/2005
5-29
Operation
D = A – B – C
with:
• A = Indication (in dB)
Absolute instrument gain with which the maximum
flaw echo is at 50 % (±5 %) echo height
• B = Reference (in dB)
Absolute instrument gain with which the maximum
reference echo (1.5 mm side-drilled hole from the
reference block 1) is at 50 % (±5 %) echo height
•
C = Attenuation (in dB)
This value is calculated according to the formula
C = 0.079 dB/mm (s – 25.4 mm). With s = sound
path of the flaw echo.
The sound attenuation correction is automatically
calculated and displayed by the instrument. For
sound paths smaller than or equal to 25.4 mm
(1 inch), the value is set to zero.
• D = D 1.1 Rating (in dB)
This is the result of the evaluation according to AWS.
The evaluation is carried out in the USM 35X accord-
ing to the formula indicated above.
Classification of welds (function group AWS)
H
Notes:
Make sure that all instrument options for the special
test are calibrated before starting with the rating ac-
cording to AWS.
Pay attention to peaking an echo with an amplitude
between 45 % and 55 % screen height. A rating is not
possible with other amplitudes.
– Apply couplant, and couple the probe to the refer-
ence block 1. Peak the echo from the 1.5 mm side-
drilled hole.
– Select the function
aSTART
, and set up the A gate
on the reference echo.
– Vary the gain so that the reference echo is displayed
at 50 % screen height.
– Choose the function
REFRNCE
, and confirm the
choice in order to save the reference gain.
Содержание Krautkramer USM 35X
Страница 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Страница 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 21: ...Krautkramer USM 35X Issue 01 04 2005 2 1 Standard package and accessories 2 Standard package and accessories 2 ...
Страница 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Страница 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Страница 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Страница 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Страница 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Страница 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 149: ...Krautkramer USM 35X Issue 01 04 2005 8 1 Interfaces and peripherals 8 Interfaces and peripherals 8 ...
Страница 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Страница 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Страница 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...