Krautkramer USM 35X
Issue 01, 04/2005
5-7
Operation
Adjusting the display range (function group BASE)
Frequency range
Adjustment range (c = 5920 m/s)
0.2 to 1 MHz
0.5 to 9999 mm/0.02" to 390"
0.5 to 4 MHz
0.5 to 9999 mm/0.02" to 390"
0.8 to 8 MHz
0.5 to 1420 mm/0.02" to 50"
2 to 20 MHz
0.5 to 1420 mm/0.02" to 50"
– Select the function
RANGE
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.
MTLVEL (Sound velocity)
Use
MTLVEL
to set the sound velocity within the test
object. You can use sound velocities between 1000 and
15000 m/s.
• Coarse adjustment, in steps as follows (m/s):
15000
9000
5000
2000
14000
8000
4000
1600
13000
7000
3250
1450
12000
6320
3130
1000
11000
6000
3000
10000
5920
2730
• Fine adjustment
1000 ... 15000 in steps of 1 m/s
A
Attention:
Always ensure that the function
MTLVEL
is correctly
set. The USM 35X calculates all range and distance
indications on the basis of the value adjusted here.
– Select the function
MTLVEL
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.
D-DELAY (Display starting point)
Here you can choose whether to display the adjusted
range (for example 250 mm) starting from the surface
of the test object, or in a section of the test object
starting at a later point. This allows you to shift the
complete screen display and consequently also the
display zero.
Krautkramer USM 35X
Issue 01, 04/2005
5-7
Operation
Adjusting the display range (function group BASE)
Frequency range
Adjustment range (c = 5920 m/s)
0.2 to 1 MHz
0.5 to 9999 mm/0.02" to 390"
0.5 to 4 MHz
0.5 to 9999 mm/0.02" to 390"
0.8 to 8 MHz
0.5 to 1420 mm/0.02" to 50"
2 to 20 MHz
0.5 to 1420 mm/0.02" to 50"
– Select the function
RANGE
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.
MTLVEL (Sound velocity)
Use
MTLVEL
to set the sound velocity within the test
object. You can use sound velocities between 1000 and
15000 m/s.
• Coarse adjustment, in steps as follows (m/s):
15000
9000
5000
2000
14000
8000
4000
1600
13000
7000
3250
1450
12000
6320
3130
1000
11000
6000
3000
10000
5920
2730
• Fine adjustment
1000 ... 15000 in steps of 1 m/s
A
Attention:
Always ensure that the function
MTLVEL
is correctly
set. The USM 35X calculates all range and distance
indications on the basis of the value adjusted here.
– Select the function
MTLVEL
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.
D-DELAY (Display starting point)
Here you can choose whether to display the adjusted
range (for example 250 mm) starting from the surface
of the test object, or in a section of the test object
starting at a later point. This allows you to shift the
complete screen display and consequently also the
display zero.
Содержание Krautkramer USM 35X
Страница 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Страница 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 21: ...Krautkramer USM 35X Issue 01 04 2005 2 1 Standard package and accessories 2 Standard package and accessories 2 ...
Страница 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Страница 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Страница 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Страница 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Страница 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Страница 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 149: ...Krautkramer USM 35X Issue 01 04 2005 8 1 Interfaces and peripherals 8 Interfaces and peripherals 8 ...
Страница 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Страница 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Страница 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...