5-18
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Calibrating the USM 35X
Calibration with straight- and
angle-beam probes
Case A: With known material velocity
Calibration process
– Set the known material velocity in
MTLVEL
(function
group
BASE
).
– Couple the probe to the calibration block.
– Set the required display range in
RANGE
(function
group
BASE
). The calibration echo must be dis-
played on the screen.
– Position the gate on one of the calibration echoes
until the sound path of the echo is indicated in the
measurement line.
– After this, change the adjustment of the function
P-DELAY
(function group
BASE
) until the correct
sound path for the selected calibration echo is
indicated in the measurement line.
Example:
You are carrying out the calibration for the calibration
range of 100 mm/5" via the function group
BASE
using
the calibration block V1 (thickness 25 mm/1") which is
laid flatwise.
– Set
RANGE
to 100 mm/5".
– Set the known material velocity of 5920 m/s
(233 "/ms) in
MTLVEL
.
– Set the gate so that it is positioned on the first
calibration echo (from 25 mm/1").
– Read the sound path in the measurement line. If this
value is not equal to 25 mm/1", change the adjust-
ment for the function
P-DELAY
until it is at 25 mm/1".
This completes the calibration of the USM 35X to the
material velocity of 5920 m/s (233 "/ms) with a calibra-
tion range of 100 mm/5" for the probe used.
5-18
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Calibrating the USM 35X
Calibration with straight- and
angle-beam probes
Case A: With known material velocity
Calibration process
– Set the known material velocity in
MTLVEL
(function
group
BASE
).
– Couple the probe to the calibration block.
– Set the required display range in
RANGE
(function
group
BASE
). The calibration echo must be dis-
played on the screen.
– Position the gate on one of the calibration echoes
until the sound path of the echo is indicated in the
measurement line.
– After this, change the adjustment of the function
P-DELAY
(function group
BASE
) until the correct
sound path for the selected calibration echo is
indicated in the measurement line.
Example:
You are carrying out the calibration for the calibration
range of 100 mm/5" via the function group
BASE
using
the calibration block V1 (thickness 25 mm/1") which is
laid flatwise.
– Set
RANGE
to 100 mm/5".
– Set the known material velocity of 5920 m/s
(233 "/ms) in
MTLVEL
.
– Set the gate so that it is positioned on the first
calibration echo (from 25 mm/1").
– Read the sound path in the measurement line. If this
value is not equal to 25 mm/1", change the adjust-
ment for the function
P-DELAY
until it is at 25 mm/1".
This completes the calibration of the USM 35X to the
material velocity of 5920 m/s (233 "/ms) with a calibra-
tion range of 100 mm/5" for the probe used.
Содержание Krautkramer USM 35X
Страница 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Страница 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 21: ...Krautkramer USM 35X Issue 01 04 2005 2 1 Standard package and accessories 2 Standard package and accessories 2 ...
Страница 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Страница 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Страница 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Страница 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Страница 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Страница 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 149: ...Krautkramer USM 35X Issue 01 04 2005 8 1 Interfaces and peripherals 8 Interfaces and peripherals 8 ...
Страница 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Страница 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Страница 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...