![GE Krautkramer USM 35X Скачать руководство пользователя страница 113](http://html.mh-extra.com/html/ge/krautkramer-usm-35x/krautkramer-usm-35x_operators-manual_128397113.webp)
Krautkramer USM 35X
Issue 01, 04/2005
5-65
Operation
OFFSET (Distance of multiple DAC)
You can activate a multiple DAC and at the same time
determine the distance from the registration curve. The
default setting 6.0 dB generates four other curves at
–12 dB, –6 dB, +6 dB, and +12 dB from the registration
curve. The setting
0
generates only the registration
curve. Any setting different from 0 generates four other
curves at a set distance from the original curve. For a
better distinction in multiple DACs the registration curve
is displayed as a bold line.
Adjustment range: 0 dB ... 14 dB in steps of 0.5 dB
– Select the function
OFFSET
.
– Use the right-hand rotary knob to select the required
setting.
Distance-amplitude curve (only USM 35X DAC and USM 35S)
Echo evaluation with DAC
In order to be able to evaluate a flaw indication by
means of the DAC, certain conditions must be met:
• The distance-amplitude curve must already be
recorded.
• It only applies to the same probe that was used
when recording the curve. Not even another probe of
the same type must be used!
• The DAC only apply to the material corresponding to
the material of the reference block.
• All functions affecting the echo amplitude must be
set the same way as they were when the curve was
recorded. This applies in particular to the following
functions:
POWER
,
FREQU
,
RECTIFY
,
MTLVEL
and
REJECT
.
Krautkramer USM 35X
Issue 01, 04/2005
5-65
Operation
OFFSET (Distance of multiple DAC)
You can activate a multiple DAC and at the same time
determine the distance from the registration curve. The
default setting 6.0 dB generates four other curves at
–12 dB, –6 dB, +6 dB, and +12 dB from the registration
curve. The setting
0
generates only the registration
curve. Any setting different from 0 generates four other
curves at a set distance from the original curve. For a
better distinction in multiple DACs the registration curve
is displayed as a bold line.
Adjustment range: 0 dB ... 14 dB in steps of 0.5 dB
– Select the function
OFFSET
.
– Use the right-hand rotary knob to select the required
setting.
Distance-amplitude curve (only USM 35X DAC and USM 35S)
Echo evaluation with DAC
In order to be able to evaluate a flaw indication by
means of the DAC, certain conditions must be met:
• The distance-amplitude curve must already be
recorded.
• It only applies to the same probe that was used
when recording the curve. Not even another probe of
the same type must be used!
• The DAC only apply to the material corresponding to
the material of the reference block.
• All functions affecting the echo amplitude must be
set the same way as they were when the curve was
recorded. This applies in particular to the following
functions:
POWER
,
FREQU
,
RECTIFY
,
MTLVEL
and
REJECT
.
Содержание Krautkramer USM 35X
Страница 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Страница 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 21: ...Krautkramer USM 35X Issue 01 04 2005 2 1 Standard package and accessories 2 Standard package and accessories 2 ...
Страница 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Страница 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Страница 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Страница 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Страница 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Страница 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 149: ...Krautkramer USM 35X Issue 01 04 2005 8 1 Interfaces and peripherals 8 Interfaces and peripherals 8 ...
Страница 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Страница 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Страница 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Страница 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...