PACE Pressure Indicators User Manual
[EN] English
6 - 25
K0470 Revision A
d.
Press the SELECT FROM RECENT button at the top of the screen to view
the IP address history.
Head Correction
Corrects pressure reading for the height difference
between instrument reference level and UUT. For
accuracy, head correction must be enabled and the
parameters set for each sensor:
•
for UUT positioned higher than the reference level
of the PACE1000 enter a positive height correction.
•
for UUT positioned lower than the reference level of the PACE1000 enter a negative
height correction.