3 Operation
K0470 Revision A
3 - 10
[EN] English
Leak testing option
This task, measures the leak rate over the measure dwell time.
At the start of the test, the instrument measures the test pressure of the user system. The
instrument then records the pressure change during measure dwell time.
On completion, the display shows the leak rate results with leak rate per second or per minute in
the current pressure units selected in measure set-up.
Start
Test time
Leak test
Settings
Leak rate
/second
/minute
Содержание K0470