6 Reference and Specification
K0470 Revision A
6 - 10
[EN] English
RS232
Located on the rear panel an external RS232 connection requires the following:
** Selectable through the user interface.
RS232 Comms Range Setup
Note:
The RS232 Comms Range values are set up by the user because of the hardware the
user desires to operate with the PACE 1000 equipment. This procedure only allows
supervisor to select these user pre-installed values. To initially define / change or
delete the RS232 Comms Range values, refer to (Ref: K0472 PACE SCPI Remote
Communications Manual).
Connector
9-way ‘D’ female wired as per Table 2-1
Communications
RS232 point-to-point only (daisy chain is not supported)
Baud Rate power-up default 9600, no parity & handshake = xon/xoff
Baud rates selectable **
2400, 4800, 9600, 19k2, 38k4, 57k6 & 115k2
Parity
None, Odd & Even
Flow control
None, Hardware & xon/xoff
Protocols
PACE SCPI
Heritage emulation
DPI 142/150, DPI 141
Terminator
CR or LF or CR/LF