eBike III basic / comfort
- 97 -
e
rror
Code
BIts
Bpu (B
lood
pressure
Module
)
Bit
Error
Reaction
Display / Info
Troubleshooting
0
not used
1
cuff loose
Warning
- check correct placement of cuff
at patient (tight)
- check cuff
2
Micro position
Warning
- check correct placement of
microphone (see User Manual)
- check microphone in cuff
- replace cuff
3
Leakage
Warning
- check cuff
- check hoses, hose connections
- check compensation container
- replace pneumatic unit
4
Errror Calibration
Module is switched off
Internal error
- perform NIBP test
- perform NIBP calibration
- replace BPU PCB
5
Error offset
Module is switched off
Internal error
- perform NIBP test
- perform NIBP calibration
- replace BPU PCB
6
not used
7
not used
8
Safety state mode
Module is switched off
Internal error
- restart eBike (OFF / ON)
- if error persists:
replace BPU
9
Setup not ready
Module is switched off
Internal error
- restart eBike (OFF / ON)
- if error persists:
check connection LIU <-> BPU
- replace BPU
10
CAN zykl. Timeout
Module is switched off
Internal error
- restart eBike (OFF / ON)
- if error persists:
check cable DTU <-> LIU
check connection LIU <-> BPU
- replace BPU