2018111-336 Rev. B
eBike III basic / comfort
- 17 -
t
est
proCedure
“B
lood
pressure
(nIBp)“
t
est
preparatIon
1. Power eBike III on and enter service menu / NIBP test (-see „Calibration of Blood pressure unit“ on page 75).
2. Connect the calibration equipment with the eBike III cuff connector.
l
eak
-
tIgHtness
test
1. Use the connected test equipment to apply the target pressure (see table below) to the eBike III cuff connector, wait
until the pressure in the system has stabilized (approx. 5 sec).
2. After 1 min check the actual pressure and note the pressure drop in the table
a
CCuraCy
test
1. Use the connected test equipment to apply the target pressure (see table below) to the eBike III cuff connector, wait
until the pressure in the system has stabilized (approx. 5 sec).
2. Check the actual pressure displayed on the eBike III display and note the value in the table.
Note
• A calibrated NIBP measure instrument set is needed to perfom the following test procedures
• Please refer to the operator manual of the used device
Note
• Unless regulatory requirements of the country define a different test procedure or test interval, it is recommended to perform the
test “Blood pressure (NIBP)“ at least every 2 years
Step
Target Pressure
Actual pressure drop
Pressure drop max.
Pass / Fail
#1
50 mmHg
mmHg / min
< 6 mmHg / min
#2
300 mmHg
mmHg / min
< 6 mmHg / min
Step
Target pressure
Displayed pressure (eBike III)
Deviation max.
Pass / Fail
#1
0
± 3 mmHg
#2
50
± 3 mmHg
#3
100
± 3 mmHg
#4
150
± 3 mmHg
#5
200
± 3 mmHg
#6
250
± 3 mmHg
#7
300
± 3 mmHg