![GE eBike III basic Series Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/ge/ebike-iii-basic-series/ebike-iii-basic-series_service-manual_131963071.webp)
2018111-336 Rev. B
eBike III basic / comfort
- 71 -
OUTPUT 3 Systole
GAIN
1000 mV
GAIN
100 mmHg
EXIT
OUTPUT 4 Diastole
GAIN
1000 mV
GAIN
100 mmHg
EXIT
OUTPUT 5 rpm
GAIN
1000 mV
GAIN
50 rpm
EXIT
a
lIgnMent
“o
utput
: a
nalog
s
ystole
“
The output voltage representing the systolic value is
defined with this setting.
It is possible to adjust the voltage and / or the systolic
value.
The setting is performed as described previously under
Alignment “Ouput: Analog Load value“
Setting range:
GAIN
(voltage):
500 to 2000 mV in
2
mV
steps
GAIN
(systole):
50 to 250 mmHg in
5
mmHg
steps
a
lIgnMent
“o
utput
: a
nalog
d
Iastole
“
The output voltage representing the diastolic value is
defined with this setting.
It is possible to adjust the voltage and / or the diastolic
value.
The setting is performed as described previously under
Alignment “Ouput: Analog Load value“
Setting range:
GAIN
(voltage):
500 to 2000 mV in
2
mV
steps
GAIN
(diastole):
50 to 250 mmHg in
5
mmHg
steps
a
lIgnMent
„o
utput
: rpM“
The output voltage representing the actual rpm value is
defined with this setting.
It is possible to adjust the voltage and / or the rpm value.
The setting is performed as described previously under
Alignment “Ouput: Analog Load value“
Setting range:
GAIN
(voltage):
200 to 1000 mV in
2
mV
steps
GAIN
(rpm):
30 to 130 rpm in
5
rpm
steps
Содержание eBike III basic Series