- 68 -
eBike III basic / comfort
2018111-336 Rev. B
Analog Calibration
Input 1 Load
Output 1
Load
Output 2
Heart.
Output 3
Syst.
Output 4
Diast.
EXIT
nIBp t
est
/ nIBp C
alIBratIon
This menu is enabled, if the ergometer is equipped with an
automatic NIBP module and this module is activated.
Test and calibration of the NIBP module is described in
session „Calibration of Blood pressure unit“ on page 75..
a
nalog
C
alIBratIon
This function is used to allign the analog remote signals
(input and output) to the corresponding ECG devices.
The eBike III ergometers have to be equipped with the
optional „COM Module (PN 201118-340)“. This module is
connected with a small cable to the 9-pin DSUB connector
and fixed to the chassis with a plastic velcro tape.
In addition, a jummper connector has to be installed on
the LIU board.
The COM module handles one analog input and five analog
output signals from/to external devices.
Select the analog signal to be aligned and confirm
by pressing the NIBP key.
a
lignment
of
analog
input
/
output
SignalS
Note
• To align the analog load signals, the corresponding ECG
recorder should be connected as a reference voltage
source
• Disconnect the RS232 or USB cable, if connected
• The alignment of the analog signals can be controlled
after exiting the service menu (e g by starting a remote
controlled stress test)
S
ignal
aSSignment
of
analog
connector
(com
module
)
Pin 1 :
INPUT Load
Pin 2:
reserved - do not connect
Pin 3:
reserved - do not connect
Pin 4:
OUTPUT BP heartrate
Pin 5:
OUTPUT Load
Pin 6:
OUTPUT Diastole
Pin 7:
OUTPUT rpm
Pin 8:
OUTPUT Systole
Housing:
Signal GND