4 Command and Query Summary
K0472 Revision A
4-22
:INST:SENS:FULL
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INSTrument:SENSe[x]:FULLscale?
where: x = 1, 2…8 is the sensor number. It defaults to 1.
Short form:
:INST:SENS:FULL?
Function:
Queries sensor full-scale value.
Response:
Returns the full-scale value of the selected sensor.
Description
This query returns the following:
x SensorPACE5000, PACE6000
PACE1000
x SensorPACE5000, PACE6000
PACE1000
1 Module 1: control sensor
Internal
sensor1
5 Module 2: control sensor
External
IDOS-2
2 Module 1: source prve
Internal
sensor 2
6 Module 2: source prve
External
IDOS-3
3 Module 1: source pressure -ve
Internal
sensor 3
7 Module 2: source pressure -ve
-
4
Module 1: barometric range
(optional)
External
IDOS-1
8
Module 2: barometric range
(optional)
-
:INST:SENS:FULL?
:INST:SENS:FULL 7.0000000
Default - module 1 control sensor full-scale.
:INST:SENS5:FULL?
:INST:SENS5:FULL 3.5000000
Returns module 2 control sensor full-scale.
:INST:SENS1:FULL?
:INST:SENS:FULL 7.0000000
Returns module 1 control sensor full-scale.
:INST:SENS6:FULL?
:INST:SENS2:FULL 20.0000000
Returns module 2 +ve source sensor full-scale.
:INST:SENS2:FULL?
:INST:SENS2:FULL 20.0000000
Returns module 1 +ve source sensor full-scale.
:INST:SENS7:FULL?
:INST:SENS7:FULL 10.0000000
Returns module 2 -ve source sensor full-scale.
:INST:SENS3:FULL?
:INST:SENS3:FULL 2.0000000
Returns module 1 -ve source sensor full-scale.
:INST:SENS8:FULL?
:INST:SENS8:FULL 1.1500000
Returns module 2 barometric sensor full-scale.
:INST:SENS4:FULL?
:INST:SENS4:FULL 1.1500000
Returns module 1 barometric sensor full-scale.