![GE Druck PACE1000 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/ge/druck-pace1000/druck-pace1000_technical-manual_132671043.webp)
PACE Series SCPI Manual
K0472 Revision A
4-15
:INST:CONT:SENS
Applicability:
PACE5000 and PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INSTrument(x):CONTroller(y]:SENse(z)
:Inst(x):Cont(y):Sens(z)?
INST
index(x)
where
x
must
be
0
or
‘blank’
Use
the
CONT
index(y)
to
determine
which
control
module
1
(or
blank)
or
2.
Both
:Inst:Cont
and
:Inst:Cont1
will
address
control
module
1,
whilst
:Inst:Cont2
addresses
only
control
module
2.
Use
the
SENS
index(z)
to
determine
which
sensor
as
below.
Description
:INST:CONT:SENS
or
:INST:CONT:SENS1
"2.00barg"
returns
Control
module
1
full
‐
scale
range
sensor
:INST:CONT:SENS2
"7.00barg"
returns
the
+ve
port
source
pressure
sensor
full
‐
scale
:INST:CONT:SENS3
"1.00barg"
returns
the
–ve
port
vacuum
pressure
sensor
full
‐
scale
:INST:CONT:SENS4
"BAROMETER"
:INST:CONT:SENS5
""
Reference
sensor
range
if
fitted
:INST:CONT:SENS6
""
Not
used
:INST:CONT:SENS7
"3.00bara"
Returns
the
pseudo
absolute
range
:INST:CONT:SENS8 "" Not used
Example
TX> :Inst:Cont1:Sens2?
RX> :INST:CONT1:SENS2 "7.00barg"
1st control module vacuum sensor
TX> :Inst:Cont2:Sens4?
RX> :INST:CONT2:SENS4 "BAROMETER"
2nd control module barometric sensor