PACE Series SCPI Manual
K0472 Revision A
4-9
:CAL:PRES:ZERO:TIME
Applicability:
PACE5000, PACE6000
Command Syntax
:CALibration[x]:[PRESsure]:ZERO:TIME
where: x (module) = 1 or 2 (default - 1)
Parameter:
Numeric
Short form:
:CAL:ZERO:TIME
Function:
Sets timed zero in hours.
Default:
-
Query Syntax
:CALibration[x]:[PRESsure]:ZERO:TIME?
where: x (module) = 1 or 2 (default - 1)
Short form:
:CAL:ZERO:TIME?
Function:
Queries timed zero.
Response:
Integer number in hours.
Description
This command sets the time between zeroing function. The query sends the setting in hours.