4 Command and Query Summary
K0472 Revision A
4-128
:UNIT:PRES:DEF
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
:UNIT[x][:PRESsure]:DEFine[y] <string>,<number>
where: x = 1 or 2 is the module number (default - 1) and y = 1, 2, 3 or 4 is the number of user
defined units (default - 1).
Parameter:
<string>
ASCII representation of unit name.
<number> Conversion factor from Pascals to required units
Short form:
:UNIT[x]DEF[y] <string>,<number>
Function:
There are four user defined units on the instrument. This command
defines the name and conversion factor to use.
minimum string
“”
number
0.0
maximum string
8 characters
number
1.0e10
Default:
string
N/A
number
1000.0
Query Syntax
:UNIT[x][:PRESsure]:DEFine[y]?
where x = 1 or 2 is the module number (default - 1) and y = 1,
2, 3 or 4 is the number of user defined units (default - 1).
Short form:
:UNIT[x]DEF[y]?
Function:
Query conversion factor
Response:
A string representing name and number
corresponding to the conversion factor.
Description
This command defines the name and conversion factor for the special units. This conversion
factor is from Pascals to the required units. An index of 1 is for USER1 unit and an index of 2
is for USER2 unit.
1
USER1
2
USER2
3
USER3
4
USER4