PACE Series SCPI Manual
K0472 Revision A
4-119
:SYST:COMM:GPIB:SELF:ADDR
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
Parameter:
Integer
Short form:
SYST:COMM:GPIB:SELF:ADDR
Function:
Sets the instrument’s GPIB address.
Query Syntax
:SYSTem:COMMunicate:GPIB:SELF:ADDRess?
Short form:
:SYST:COMM:GPIB:SELF:ADDR?
Function:
Queries GPIB address
Response: Returns
decimal
number representing GPIB address.
Description
This command instructs the instrument to set a GPIB address. The query command gets the
current address.
Example:
:TX> :SYST:COMM:GPIB:SELF:ADDR?
RX> :SYST:COMM:GPIB:SELF:ADDR 1
//Current GPIB address is 1
TX> :SYST:COMM:GPIB:SELF:ADDR 16
//Set address to 16
TX> :SYST:COMM:GPIB:SELF:ADDR?
RX> :SYST:COMM:GPIB:SELF:ADDR 16