K114 Issue No. 9
6 - 9
Druck ADTS 405 User Manual
6
PORT
This function sets the port option, this changes the ADTS 405 from a single pitot-static
output system to a multi-port system by using a Druck Line Switching Unit (LSU 100
and 200 Series). Refer to the option manual.
REMOTE
Enters Remote Mode; this mode allows an external device to control the unit. In this
mode the key-pad (except the Remote key) is disabled. If remote lock is disabled,
pressing the remote key when the
"REMOTE"
message is not displayed forces the unit
into remote mode as if the external device has taken control of the unit.
Note:
This may change operator settings such as the current limit set.
Refer to the option user manual for further details.
The ADTS 405 contains a print facility to print the displayed readings together with up
to nine user readings (e.g., Pilot, Copilot, Aux).
This facility requires a Centronics standard parallel interface printer of 80 or 132
column width.
This key causes a printout of the date and time followed by the measured value of the
parameters. A prompt asks for the entry of a reading which can consist of any
characters
A
to
Z
,
0
to
9
,
.
(decimal point) or
-
(minus), in any order. It can be used to
enter the value from the unit or system under test or alternatively a test number.
When no more entries are required, press
CLEAR/QUIT
.
To enter an alphabetic character (A to Z), press the [ALPHA] key and use the
s
or
t
(nudge keys) to change to the required character. To enter another alphabetic
character, press the [ALPHA] key again. Press [BACK] to delete the last character.
Up to nine user readings can be entered sequentially. The display prompts for the next
user reading each time one is entered. When no further readings are required, press
CLEAR/QUIT
.
Example print output:
04Mar00 09:47 Ps 700.00 mbar Qc 30.00 mbar Usr Rdg
Содержание Druck ADTS 405
Страница 2: ... General Electric Company All rights reserved ...
Страница 4: ...K114 Issue No 9 ...
Страница 32: ...K114 Issue No 9 2 10 Druck ADTS 405 User Manual F IGURE 2 4 ADTS 405 G ENERAL VIEW ...
Страница 33: ...K114 Issue No 9 Installation 2 11 2 F IGURE 2 5 ADTS 405 R EAR P ANEL VIEW ...
Страница 41: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 7 3 Control Mode ...
Страница 43: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 9 3 Go to Ground ...
Страница 50: ...K114 Issue No 9 3 16 Operation FIGURE 3 6 TRIPLE DISPLAY FIGURE 3 5 DUAL DISPLAY ...
Страница 72: ...K114 Issue No 9 4 6 Maintenance INTENTIONALLY LEFT BLANK ...
Страница 92: ...K114 Issue No 9 5 20 Testing and Fault Finding INTENTIONALLY LEFT BLANK ...