4
Cable tests
Carry out the following check when detailed in Testing and Fault Finding, Section 5.
Measure continuity using the DVM (item 1, Table 4-3) set to an appropriate range.
Measure the continuity between corresponding pins at each end of each cable
assembly. The measured resistance must not exceed 2.0 ohms.
Note
: This resistance value is the expected maximum value for the 18 metre (60ft) cable.
The resistance value varies depending on the length of the cable.
Measure insulation using the Insulation tester (item 2, Table 4-3), set to an appropriate
range.
Measure the insulation resistance between the shell of the connector and all individual
pins in turn.
Measure the insulation resistance between the shell of the appropriate connectors and
all pins in turn.
Measure also the resistance between individual pairs of pins (i.e., 1-2, 1-3, 1-4, 1-5, 1-6,
2-3, 2-4, 2-5, 2-6, 3-4, 3-5, 3-6, 4-5, 4-6 and 5-6). In all cases the resistance must
exceed 10 Mohms.
Содержание Druck ADTS 405
Страница 2: ... General Electric Company All rights reserved ...
Страница 4: ...K114 Issue No 9 ...
Страница 32: ...K114 Issue No 9 2 10 Druck ADTS 405 User Manual F IGURE 2 4 ADTS 405 G ENERAL VIEW ...
Страница 33: ...K114 Issue No 9 Installation 2 11 2 F IGURE 2 5 ADTS 405 R EAR P ANEL VIEW ...
Страница 41: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 7 3 Control Mode ...
Страница 43: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 9 3 Go to Ground ...
Страница 50: ...K114 Issue No 9 3 16 Operation FIGURE 3 6 TRIPLE DISPLAY FIGURE 3 5 DUAL DISPLAY ...
Страница 72: ...K114 Issue No 9 4 6 Maintenance INTENTIONALLY LEFT BLANK ...
Страница 92: ...K114 Issue No 9 5 20 Testing and Fault Finding INTENTIONALLY LEFT BLANK ...