Programming Site Data
1-35
April 2008
Setting the A Value
In the Skan/Measure Integrate mode, the A value aids in minimizing
cycle skipping. Use this option to specify parameters for the A value.
IMPORTANT:
The
AVAL
default settings are suitable for most
applications. Before changing any of these
parameters, consult the factory.
1.
To enter the
AVAL
submenu, press
[F2]
at the
ADVAN
window.
2.
Use the numeric keys to enter the desired
A_Value Minimum
, and
press
[ENT]
. (The available range is from 75 to 100, and the default
value is 75.)
3.
Use the numeric keys to enter the desired
A_Value Maximum
, and
press
[ENT]
. (The available range is from 100 to 125, and the
default value is 125.)
4.
Enter the desired
-Offset % of Peak
from 0 to 95%, and press
[ENT]
. (The default value is 20%.)
5.
Enter the number of
Cycle Count Samples
(from 1 to 40) and press
[ENT]
. (The default number is 8.)
6.
Enter the
Cycle #
(from 1 to 100) and press
[ENT]
. (The default
number is 1.)
The GC868 returns to the
ADVAN
window.
Содержание DigitalFlow GC868
Страница 8: ...Chapter 1 ...
Страница 61: ...Chapter 2 ...
Страница 78: ...Chapter 3 ...
Страница 91: ...Chapter 4 ...
Страница 105: ...Chapter 5 ...
Страница 106: ...Clearing Data Introduction 5 1 Reset Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 111: ...Chapter 6 ...
Страница 123: ...Appendix A ...
Страница 132: ...Appendix B ...
Страница 133: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...