Programming Site Data
1-13
April 2008
Entering Pipe Data
The
PIPE
submenu permits entry of the transducer and pipe
specifications. To program this menu, complete the following steps:
1.
At the
User Program
screen, press
[F3]
to program the
PIPE
submenu.
2.
Enter the
Transducer Number
(normally engraved on the head of
the transducer). Press
[ENT].
If there is no engraved number,
complete the steps below. Otherwise, proceed to step 3.
IMPORTANT:
Special transducers, which have no engraved
number on the head, are rarely used. Examine the
transducer head carefully for a number.
Special Transducers
a.
Assign a number between 91 and 99 to the
Special Transducer
and press
[ENT]
. (The meter will only accept values from 1 to
199.)
b.
Press
[F2]
to select Shear wave as the
Transducer Type
.
Note:
While three choices are available (Rayleigh, Shear wave or
wetted transducers), the GC868 is designed for use with Shear
wave clamp-on transducers. Consult the factory before
selecting Rayleigh or wetted transducers.
c.
Use the [
→]
and
[F1]-[F4]
keys to select the
Frequency
of the
special transducer. The meter can not transmit an excitation
voltage at the transducer’s natural frequency without this data.
d.
Press
[F1]-[F4]
to select the
Wedge Material
. The available
options include
VHT
(very high temperature),
MT
(medium
temperature),
HT
(high temperature) and
OTHER
. Refer to the
data sheet supplied with your transducer to determine the
appropriate setting.
e.
Use the numeric keys to enter the
Wedge Angle
(the angle of
the ultrasonic transmission) supplied by the factory in degrees
and press
[ENT]
.
•
If you selected
OTHER
in step d, use the numeric keys to enter
the
Wedge Soundspeed
in ft/sec or m/sec supplied by GE and
press
[ENT]
. The meter proceeds to step 5.
Wedge Temperature
3.
Use the numeric keys to enter the
Wedge Temperature
supplied by
the factory in deg F and press
[ENT]
.
Содержание DigitalFlow GC868
Страница 8: ...Chapter 1 ...
Страница 61: ...Chapter 2 ...
Страница 78: ...Chapter 3 ...
Страница 91: ...Chapter 4 ...
Страница 105: ...Chapter 5 ...
Страница 106: ...Clearing Data Introduction 5 1 Reset Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 111: ...Chapter 6 ...
Страница 123: ...Appendix A ...
Страница 132: ...Appendix B ...
Страница 133: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...