April 2008
1-10
Programming Site Data
Entering System Data
1.
At the
User Program
screen, press the
[F2]
function key to
program the
SYSTM
submenu.
2.
Enter a
Site Label
of up to 9 characters and press
[ENT]
. (While
taking measurements, the site label will appear on the locator bar.)
3.
Enter a
Site Message
of up to 21 characters. Press
[ENT]
.
Entering System Units
4.
To select the
System Units
, press
[F1]
to display parameters and
measurements in English units, or press
[F2]
to display parameters
and measurements in Metric units.
5.
Use the
[F1]
-
[F4]
keys to select the type of
Pressure Units
desired.
The available pressure units are shown in Figure A-1 on page A-1.
The choices shown on the option bar are determined by the
selections made at the previous
SYSTEM UNITS
prompt.
a.
If you have entered the local
Atmospheric Pressure
(PSIg,
BARg or kPag), use the numeric keys to enter the atmospheric
pressure value. Press
[ENT].
6.
At the
Stopwatch Totalizer
prompt, press
[F1]
to totalize all liquid
flow continuously, or
[F2]
to measure totals manually with the
Stopwatch Timer. (With
MNUAL
(
[F2])
, the console key on the
keypad is used to start and stop the totalizer. See page 1-7 for
details.)
IMPORTANT:
After setting up the Stopwatch Totalizer, press
[CLR]
to
clear the Stopwatch Totalizer, or the new totals will
be added to any previously accumulated totals.
Note:
The manner in which the Stopwatch Totalizer responds to an
error condition may be set in the
I/O
submenu.
7.
At the
Equation
prompt, press
[F1]
to display the measurement
data in
standard
volumetric units or press
[F2]
to display the
measurement data in
actual
volumetric units.The Model GC868
will use the appropriate gas equation to calculate the flow rate,
corresponding to the measured data, in the volumetric units
indicated.
a.
If you selected
Standard
, press
[F1]
to have the meter treat the
gas as an
ideal gas
or press
[F2]
to have the meter treat the gas
as a
supercompressed gas
.
Содержание DigitalFlow GC868
Страница 8: ...Chapter 1 ...
Страница 61: ...Chapter 2 ...
Страница 78: ...Chapter 3 ...
Страница 91: ...Chapter 4 ...
Страница 105: ...Chapter 5 ...
Страница 106: ...Clearing Data Introduction 5 1 Reset Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 111: ...Chapter 6 ...
Страница 123: ...Appendix A ...
Страница 132: ...Appendix B ...
Страница 133: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...