Programming Site Data
1-21
April 2008
Programming a Totalizer
Output
5.
The
TTLZR
output,
[F3]
, issues one pulse per selected volume of
flow. The meter produces a pulse each time the programmed
amount of flow passes through the pipe. Complete the steps below
to program the totalizer output.
a.
Press
[F1] = +TOTL
to totalize the forward flow,
[F2] = -TOTL
to
totalize the reverse flow,
[F3] = +MASS
(if available) to totalize
the forward mass flow or
[F4] = -MASS
(if available) to totalize
reverse mass flow.
b.
Enter a value from 50 to 500,000
μ
sec for the
Minimum Pulse
ON
Time
and press
[ENT]
.
Note:
A complete pulse consists of equal amounts of
ON
and
OFF
times. Choose a value that is compatible with the frequency
counter to be used.
c.
Enter a value for the number of measurement
Units/Pulse
and
press
[ENT]
.
6.
Press
[F1]
-
[F4]
to set up another totalizer/frequency output or press
[EXIT]
to select another slot for setup.
Содержание DigitalFlow GC868
Страница 8: ...Chapter 1 ...
Страница 61: ...Chapter 2 ...
Страница 78: ...Chapter 3 ...
Страница 91: ...Chapter 4 ...
Страница 105: ...Chapter 5 ...
Страница 106: ...Clearing Data Introduction 5 1 Reset Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 111: ...Chapter 6 ...
Страница 123: ...Appendix A ...
Страница 132: ...Appendix B ...
Страница 133: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...