Programming Site Data
1-51
March 2008
Programming a Totalizer
Output
This type of output issues one pulse per selected volume of flow. The
meter produces a pulse each time the programmed amount of flow
passes through the pipe.
9.
Press
[F1]-[F4]
to select the parameter to be totalized. See Table 1-
24 below for a description of the available options.
Note:
The measurement units that appear in these prompts are those
selected in the
GLOBL-SYSTM
menu earlier in this chapter.
10.
Enter a value from 50 to 4,000,000
μ
sec for the
Minimum
ON
Time
of each totalizer pulse and press
[ENT]
.
Note:
A complete pulse consists of equal amounts of
ON
and
OFF
times. Choose a value that is compatible with the frequency
counter to be used.
11.
Enter a value for the number of measurement
Units
represented by
each pulse and press
[ENT]
.
12.
Press
[F1]
-
[F4]
to set up another totalizer/frequency output or press
[EXIT]
to select another slot for set up.
Table 1-24: Output Measurement Options
Option Bar Choice
Description
[F1]
=
+TOTL
Forward Totalized Volume Flow
[F2]
=
-TOTL
Reverse Totalized Volume Flow
[F3]
=
+MASS*
Forward Totalized Mass Flow
[F4]
=
-MASS*
Reverse Totalized Mass Flow
[
→]
+ [F1]
= +ENRG**
Forward Energy Flow
[
→]
+ [F2]
= -ENRG**
Reverse Energy Flow
*Available only if
Mass Flow (static)?
=
YES
**
Available only if
Energy Option = YES
Содержание DigitalFlow DF868
Страница 7: ...Chapter 1 ...
Страница 75: ...Chapter 2 ...
Страница 91: ...Chapter 3 ...
Страница 105: ...Chapter 4 ...
Страница 116: ...Chapter 5 ...
Страница 117: ...Clearing Data Introduction 5 1 Clearing Totalized Data 5 2 Deleting Site Files 5 2 Deleting Log Files 5 3 ...
Страница 121: ...Chapter 6 ...
Страница 132: ...Appendix A ...
Страница 142: ...Appendix B ...
Страница 143: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 149: ...Appendix C ...
Страница 169: ...Appendix D ...